UVT-S-AR DNA/RNA UV-cleaner box

DNA/RNA UV-cleaner box UVT-S-AR is designed for clean operations with DNA samples. UV-cleaner box provide protection against contamination.

Model is a bench-top type, made of metal framework, glass walls, working surface made of stainless steel.

UV-cleaner boxes are equipped with an open UV lamp installed in the upper hood. UV-radiation from the open lamps disinfects the working area inactivating DNA/RNA fragments during 15–30 min of exposure. A digital timer controls duration of the direct UV irradiation. A daylight lamp provides proper illumination of the working surface.

UV-cleaner boxes are equipped with a flow-type bactericidal UV cleaner–recirculator AR, which provides constant decontamination inside the box during operation. They are recommended for operations with DNA/RNA amplicons.

UV cleaner–recirculator AR consists of a UV lamp, a fan and dust filters organized in a special body so that a user working with a UV-cleaner box is protected against UV light. Recirculator increases the maximum density of UV light making it  sufficiently effective for DNA/RNA inactivation. The UV–recirculator processes 100 UV-cleaner box volumes per hour, creating permanent aseptic conditions of operation inside the UV-cleaner box.

Specially assigned moving tables T-4L (with wheel locks) with a drawer are available on request.

Advantages of Biosan UV-cleaner boxes:

  • Ozone free high density UV decontamination
  • Long living UV lamps (9,000 hours average)
  • Automatic switch off of UV-lamps when the protective screen is opened
  • Bactericidal flow-type recirculator providing permanent decontamination inside UV – cleaner box during operation
  • Shockproof glass walls
  • Low noise, low energy consumption
  • Tables for installation of UV-cleaner boxes
  • UV-cleaner boxes with the bactericidal UV cleaner–recirculator AR is the patented Biosan solution

 

Specification

Walls material

Rear: stainless steel Sides and

front: glass (EUROGLASS, Germany)

Working surface material

Stainless steel

Open UV-lamp

2x30W built‐in bactericidal, TUV30WG13 UV-C

UV radiation level

15 mW / cm2 / s

Radiation type

UV (λ = 253.7 nm), ozone-free

Digital time setting of

1 min–24 hrs / non–stop

direct UV exposure

(increment 1 min)

UV-recirculator

1 х 30W (efficiency >99% per 1 hour)

Daylight lamp (for working area

illumination)

1 x TLD-30W

Thickness of side panels

4 mm

Thickness of upper front panel

8 mm

Thickness of screen

5 mm

Optical transmission

95%

UV protection

>96%, UV-protection film

Working area

1200 × 520 mm

Safety features

Automatic open UV-lamp switch off when screen is open

Power outlets inside

the unit

3 Built-in sockets max. 1000 W

and inlet for power cords

Overall dimensions (W×D×H)

1245 × 585 × 585 mm

Weight (net / gross)

58 / 68.5 kg

Power consumption

135 W

Nominal operating voltage

100-240 V, 50/60Hz

Optional table

T-4L (1290 × 600 × 770 mm)

Optional drawer unit

LF-1 (300 × 450 × 705 mm)