DNA/RNA UV-cleaner box UVC/T-M-AR is designed for clean operations with DNA samples. UV--cleaner box provide protection against contamination.
Model is a bench-top type, made of metal framework, glass walls, working surface made of stainless steel.
UV-cleaner boxes are equipped with an open UV lamp installed in the upper hood. UV-radiation from the open lamps disinfects the working area inactivating DNA/RNA fragments during 15–30 min of exposure. A digital timer controls duration of the direct UV irradiation. A daylight lamp provides proper illumination of the working surface.
UV-cleaner boxes are equipped with a flow-type bactericidal UV cleaner–recirculator AR, which provides constant decontamination inside the box during operation. They are recommended for operations with DNA/RNA amplicons.
UV cleaner–recirculator AR consists of a UV lamp, a fan and dust filters organized in a special body so that a user working with a UV-cleaner box is protected against UV light. Recirculator increases the maximum density of UV light making it sufficiently effective for DNA/RNA inactivation. The UV–recirculator processes 100 UV-cleaner box volumes per hour, creating permanent aseptic conditions of operation inside the UV-cleaner box.
Specially assigned moving tables T-4 (with wheel locks) with a drawer are available on request.
Advantages of Biosan UV-cleaner boxes:
- Ozone free high density UV decontamination
- Long living UV lamps (9,000 hours average)
- Automatic switch off of UV-lamps when the protective screen is opened
- Bactericidal flow-type recirculator providing permanent decontamination inside UV –cleaner box during operation
- Shockproof glass walls
- Low noise, low energy consumption
- Tables for installation of UV-cleaner boxes
- UV-cleaner boxes with the bactericidal UV cleaner–recirculator AR is the patented Biosan solution
Specification
Walls material |
Rear: stainless steel Sides and front: glass (EUROGLASS, Germany) |
Working surface material |
Stainless steel |
Open UV-lamp |
1 x 25W built‐in bactericidal, TUV25WG13 UV-C |
UV radiation level |
15 mW / cm2 / sec |
Radiation type |
UV (λ = 253.7 nm), ozone-free |
Digital time setting of |
1 min–24 hrs / non–stop (increment |
direct UV exposure |
1 min) |
UV-recirculator |
1x25W (efficiency >99% per 1 hour) |
Daylight lamp (for working area illumination) |
1 x TLD-15W |
Thickness of side panels |
4 mm |
Thickness of upper front panel |
8 mm |
Thickness of screen |
4 mm |
Optical transmission |
95% |
UV protection |
>96%, UV-protection film |
Working area |
650 × 475 mm |
Safety features |
Automatic open UV-lamp switch off when screen is open |
Power outlets inside the unit |
Inlet for power cords or optional 1x Built-in socket, max. 1000 W (pls, order respectively) |
Overall dimensions (W×D×H) |
690 × 535 × 555 mm |
Weight (net / gross) |
28.8 / 39 kg |
Power consumption |
67 W |
Nominal operating voltage |
100-240 V, 50/60Hz |
Optional table |
T-4 (800 × 600 × 750 mm) |