UVC/T-AR DNA/RNA UV-cleaner box

DNA/RNA UV-cleaner box UVC/T-AR is designed for clean operations with DNA samples. UV-cleaner box provide protection against contamination.

Model is a bench-top type, made of metal framework, plexiglas walls and working surface painted with powder enamel.

UV-cleaner boxes are equipped with an open UV lamp installed in the upper hood. UV-radiation from the open lamps disinfects the working area inactivating DNA/RNA fragments during 15–30 min of exposure. A digital timer controls duration of the direct UV irradiation. A daylight lamp provides proper illumination of the working surface.

UV-cleaner boxes are equipped with a flow-type bactericidal UV cleaner–recirculator AR, which provides constant decontamination inside the box during operation. They are recommended for operations with DNA/RNA amplicons.

UV cleaner–recirculator AR consists of a UV lamp, a fan and dust filters organized in a special body so that a user working with a UV-cleaner box is protected against UV light. Recirculator increases the maximum density of UV light making it  sufficiently effective for DNA/RNA inactivation. The UV–recirculator processes 100 UV-cleaner box volumes per hour, creating permanent aseptic conditions of operation inside the UV-cleaner box.

Specially assigned moving tables T-4 (with wheel locks) with a drawer are available on request.

Advantages of Biosan UV-cleaner boxes:

  • Ozone free high density UV decontamination
  • Long living UV lamps (9,000 hours average)
  • Automatic switch off of UV-lamps when the protective screen is opened
  • Bactericidal flow-type recirculator providing permanent decontamination inside UV-cleaner box during operation
  • Low noise, low energy consumption
  • Tables for installation of UV-cleaner boxes
  • UV-cleaner boxes with the bactericidal UV cleaner–recirculator AR is the patented Biosan solution

 

 Specification

Walls material

Plexiglas: Polymethyl methacrylate ALTUGLAS EX

Working surface material

Steel with chemicals resistant powder coating

Open UV-lamp

1 x 25W built‐in bactericidal, TUV25WG13 UV-C

UV radiation level

15 mW/cm2 /sec

Radiation type

UV (λ = 253.7 nm), ozone-free

Digital time setting of

1 min–24 hrs / non–stop

direct UV exposure

(increment 1 min)

UV-recirculator

1x25W (efficiency >99% per 1 hour)

Daylight lamp (for working area illumination)

1 x TLD-15W

Thickness of side panels

4mm

Thickness of upper front panel

8mm

Thickness of screen

8 mm

Optical transmission

92%

UV protection

>99.90% Polymethyl methacrylate ALTUGLAS EX

Working area

650 × 475 mm

Safety features

Automatic open UV-lamp

switch off when screen is open

Power outlets inside the unit

Inlet for power cords

Overall dimensions (W×D×H)

690 × 535 × 555 mm

Weight (net / gross)

23 / 33 kg

Power consumption

67 W

Nominal operating voltage

100-240 V, 50/60Hz

Optional table

T-4 (800 × 600 × 750 mm)