Defibrillator
Defibrillator Series |
Heart+ResQ ™ NT-381.C |
Set Configuration |
Defibrillator, battery pack, sealed electrode pads, Heart+ResQ User Guide, Protective Carrying Case,"NT-MPR" CD-ROM, Bluetooth Adaptor (optional) |
Waveform |
Truncated Exponential Biphasic |
Energy |
150J (adult, impedance 500) |
ShockChargelimeI |
9 seconds (土10%) |
Shock Method |
Manual button (flashes when ready) |
Mode Conversion |
Adult/Child toggle button |
Physical
Dimensions |
220mm (W) × 282mm (L) × 80mm (H) |
Weight |
2kg (including battery and pads) |
Environmenta
Water resistance |
IP55 per IEC 60529 |
Temperature |
Operating: 0°C to 40°C |
Humidity |
10% to 95% Relative Humidity, Non-Condensing |
Altitude |
0 to 5000 meters |
Drop test |
1.5 meters per IEC 60068-2-32 |
ECG
ECG Record |
LEAD II |
Analysis Overview |
Patient evaluated to determine whether ECG rhythm represents shockable or non-shockable state. |
Analysis time |
Approx. 7 seconds (Read: about 5 seconds, Analyze: within 2 seconds) |
Data
Storeda data |
ECG recordings, operation information (whether shock delivered or not), ambient noise, self-diagnostics history |
Memory capacity |
ECG: consecutively make records during analysis and CPR (once in every 5 sec, up to 5,000 sec) |
Storage type |
Internal Memory |
Access/search |
"NT-ADV" Software (for PC) via Bluetooth |
Self-diagnostics Test
Daily self-diagnostics tests ensure that circuits for measuring patient impedance, control circuit internal discharge of electric energy, and control circuit for charge/discharge of electric energy are functioning |
Upon completion of daily self-diagnostics test, results are automatically stored within device's internal memory. |
Battery Check
Battery Reserve Level |
High: 3 bars |
Battery Status Audible Warnings |
Low: Voice prompt states |
Electrode pads
Shelf Life |
2 years, for one-time use only |
Conductive Gel |
PE foam, Hydrogel |
Conductive Element |
AgCl (Silver Chloride) |
Cable Length |
2.1m (+1%) |
Applicable Ages |
All ages (12 months & older) |
Sealing |
Airtight |
Size |
125mm x 95mm |
Pads Placement |
Anterior-lateral (visual details on pads) |
Battery pack
Type |
Non-rechargeable lithium (LiMn02) |
Voltage |
DC 12V / 4,200mAH |
Standby Mode Life |
Recommended five (5) years @ 25°C |
Capacity |
180 shocks @ 25°C, 130 shocks @ 0°C |
Dimensions |
2.2cm (H) x 16.4cm (W) x 5.4cm (D) |
Weight |
260g (士10 %) |