Miura One

Fully-automated Clinical Chemistry Analyzers up to 180 tests/hour

Capable of up to 180 tests per hour with ISE module on board, the Miura One is smart random access fully automated analyzer.

The best choice for labs, doctor offices, clinics and veterinary labs. Only 80 cm in length, the unit fits easily into most laboratory benches.

Miura One is 100% developed and produced in Italy.


Characteristics

  • up to 180 tests per hour with integrated ISE module.
  • 2 different reagents/samples configuration.
  • 19 reagents on board with 50ml R1 vials
    and/or 20ml R2 vials plus 9 samples plus diluent.
  • 19 reagents on board with 35ml vials
    and/or 15ml R2 vials plus 15 samples plus diluent.
  • Integrated ISE module for Na+, K+, Li+, Cl- in serum,
    plasma and diluted urine.
  • Integrated washing station.
  • Integrated level and shock sensor for samples
    and reagents.
  • Integrated positive identification for samples
    and reagents.
  • Integrated micro metering diluter. No glass syringe.
  • 2 years warranty.

 

TECHNICAL  SPECIFICATIONS

  • Up to 180 photometric tests per hour.
  • Analytical methods: end point mono or bichromatic, fixed time, kinetic, differential sample blank.
  • ISE module for: Na+, K+, Li+, Cl- in serum, plasma and diluted urine.
  • 19 parameters on line plus diluent with 50ml R1/ 20ml R2 vials.
  • 19 parameters on line plus diluent with 35ml R1/ 15ml R2 vials.
  • Mono, double, triple reagents method.
  • Reaction temperature: +37°C ± 0.1°C
  • 9/15 sample/calibrator/control positions on primary tubes (Ø12÷13mm × 75/85mm) and sample cups (3ml).
  • Automatic dilution of the samples.
  • Calibration with linear/non linear curves up to 8 points, auto dilution of the calibration curve (linear regression, linear piece-wise, cubic spline, polynomial).
  • Control menu with Levy-Jennings plot.
  • Bar-code reader integrated for samples and reagents (optional).
  • Arm equipped with liquid level detector (capacitive) and shock sensor.
  • High Precision Micro-metering pump to dispense reagents and samples (no syringe).
  • Reagent volume: 1÷400 microliters.
  • Sample volume: 1÷300 microliters.
  • Typical reaction volume: 200 microliters.
  • Wavelengths on-board: 8 from 340nm to 700nm plus one free position.
  • Light source: Long life halogen lamp (2000 hours).
  • Photometric linearity: 0÷3.0 OD.
  • Reading cuvette: 80 on-board re-washable made in Bionex©.
  • Cuvette washing: automatic washing station.
  • Water consumption: 2 liters per hour.
  • Reagent cooler: 14°C below ambient temperature.
  • L.I.S.: Bi-directional connection.
  • Environment temperature 18-32°C
  • Dimensions: 80x62x60 cm
  • Weight: 42 kg.